Product Focus:
|
INTENDED USE: Xpert FII & FV NOR/MUT Control is provided for Research Use Only (RUO). It cannot be cloned, sold, or transferred without the explicit written consent of MMQCI. PRODUCT SUMMARY and PRINCIPLE: Best practice is to establish a quality control program for every assay performed by the laboratory.1, 2 Routine use of quality controls that are consistent lot to lot and monitor the entire assay assists the laboratory in identifying shifts, trends, and increased frequency of random errors caused by variations in the test system, such as failing reagents. Early investigation can prevent failed assay runs. COMPOSITION: Specific mutations present are described below.
STORAGE AND STABILITY: Upon receipt and after opening, the material should be stored at 2° – 8°C. Do not freeze.Unopened Xpert FII & FV NOR/MUT Control material is stable through the expiration date printed on each bottle when stored refrigerated (2° – 8°C).Opened material tightly capped and returned to the refrigerator (2° – 8°C) shortly after use is stable for thirty (30) days from the date of opening.
INSTRUCTIONS FOR USE: Handle Xpert FII & FV NOR/MUT Control as you would a whole blood specimen:
1. Allow Xpert FII & FV NOR/MUT Control to come to room temperature (18° – 25°C). 2. Thoroughly mix the solution prior to opening by inverting the bottle several times or placing on an automated mixer immediately before use. 3. Follow manufacturer’s test instructions for clinical specimens. 4. Tightly recap each bottle after use and store refrigerated (2° ‑ 8°C). 5. Controls should be tested routinely as a matter of Good Laboratory Practice (GLP) and according to guidelines or requirements of local, state, and/or federal regulations or accrediting organizations. PRECAUTIONS AND WARNINGS: Xpert FII & FV NOR/MUT Control is for use on the GeneXpert® System only. Please contact MMQCI for controls to be used for other test systems that detect mutations Factor II G20210A and Factor V Leiden G1691A.This product contains 23% ethanol (v/v) and could be flammable. Keep away from open flames.This product is intended for in vitro analytical testing and is provided for Research Use Only, not for use in diagnostic procedures. This product does not contain any biological material of human origin.
EXPECTED VALUES:
Other Factor II and Factor V mutations are not detected in the Xpert FII & FV NOR/MUT Control.The laboratory should follow Good Laboratory Practice (GLP) and establish its own performance characteristics for the Xpert FII & FV NOR/MUT Control in demonstrating adequate system performance. ORDERING INFORMATION: Xpert FII & FV NOR/MUT Control
Order Number: G11601 Kit contains: 4 x 0.5mL bottles ( 2 Normal & 2 Mutant)1. ISO 15189: Medical laboratories – Particular requirements for quality and competence. 2. CAP Molecular Pathology Checklist; Commission on Laboratory Accreditation, Laboratory Accreditation Program, Mol.20000< |
||||||||||||||||
|
|||||||||||||||||